Important Control Solution Test Information
• Use only FORA control solutions. Check the expiration date on
the control solution vial. Do not use if expired.
• Control solution, monitor, and test strips should be at room
temperature, 68 to 77°F (20 to 25°C), before testing.
• Shake the vial before use. Discard the first drop of control solu-
tion. After squeezing, wipe off the dispenser tip to avoid contami-
nation. The above tips ensure you to get a clean sample and an
accurate result.
• Use for only 3 months after the first opening date. Record the
opening date on the control solution vial. Discard the remaining
control solution 3 months after this date.
• Store the control solution tightly closed at temperatures
36-86°F (2-30°C). Do not freeze.
The control solution range is printed on FORA test strip vial. Only use
FORA control solution with FORA test strips. It is used to evaluate moni-
tor and test strip performance. It is not the recommended range for your
blood glucose level.
24