5.2

Troubleshooting

Problem
The reading is
extremely low (or high).
Arm cuff pressure does
not rise.
Arm cuff deflates too
soon.
Cannot measure or
readings are too low or
too high.
The unit loses power
during measurement.
Cause
Arm cuff not applied
correctly.
Movement or talking
during measurement.
Clothing is interfering
with the arm cuff.
Is the air tube securely
connected into the main
unit?
Air is leaking from the
arm cuff.
The arm cuff is loose.
Has the arm cuff been
inflated sufficiently?
Is the air release button
being pressed during
inflation?
The batteries are empty.
5.Handling Errors and Problems
Remedy
Apply the arm cuff
correctly.
Refer to section 3.2.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air tube
is connected securely.
Refer to section 3.2.
Replace the arm cuff with
a new one.
Refer to Chapter 7.
Apply the cuff correctly so
that it is firmly wrapped
around the arm.
Refer to section 3.2.
Inflate the cuff so that it is
30 to 40 mmHg above
your previous
measurement result.
Refer to section 3.3.
Be careful not to press the
air release button during
measurement.
Replace the batteries with
new ones.
Refer to section 2.1.
29
EN