Appendix A: EMC Information
English
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of
this device should make sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of
this device should make sure it is used in such an environment.
Immunity Test
Electrostatic Discharge (ESD)
IEC61000-4-2
Electrical fast transient/ burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic field
IEC61000-4-8
NOTE: U
is the a.c. mains voltage prior to the application of the test level
T
Compliance
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal
Group1
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment
Class B
The device is suitable for use in all establishments,
including domestic establishments and those directly
Class A
connected to the public low-voltage power supply
network
Complies
IEC60601
test
level
±6kV contact
±8kV air
±2kV for power supply
line
±1kV for input/out line
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
<5 % U
(>95 % dip in
T
U
)for 0,5 cycle
T
40 % U
(60 % dip in
T
U
)for 5 cycles
T
70 % U
(30 % dip in
T
U
)for 25 cycles
T
<5 % U
(>95 % dip in
T
U
)for 5 sec
T
3 A/m
Compliance
±6kV contact
±8kV air
±2kV for power supply
line
±1kV for input/out line
± 1 kV line(s) to
line(s)
<5 % U
(>95 % dip in
T
U
) for 0,5 cycle
T
40 % U
(60 % dip in
T
U
) for 5 cycles
T
70 % U
(30 % dip in
T
U
) for 25 cycles
T
<5 % U
(>95 % dip in
T
U
) for 5 sec
T
3 A/m
13
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of atypical commercial or hospital
environment
Mains power quality should be that
of atypical commercial or hospital
environment.
Mains power quality should be that
of atypical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions, it
is recommended that the device be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of
atypical location in a typical
commercial or hospital
environment.