Additionally, improper filling or lower than normal operating pressure
in the Reservoir will contribute to the unit not pulsing.
3. The Portable stops pulsing during use.
• Ensure that both connections of the dual lumen cannula are firmly
attached to the Oxygen Outlet and Sensor connectors.
• Change the cannula if water droplets are present from humidified
exhaled gas. Replacing the cannula with a dry one enhances the
ability of the Portable to pulse.
• Ensure that the cannula is not kinked.
• Ensure that the cannula tips remain in your nostrils and do not
slide to one side.
• Ensure that there is oxygen in the Portable.
• With the cannula on, close your mouth and breathe only through
your nose to verify that the Portable has stopped pulsing.
4. The Portable does not fill.
• Verify that there is oxygen in the Reservoir.
• Ensure that the Portable and Reservoir fill connectors are fully
engaged throughout the filling process.
5. The Portable vent valve does not close properly at the end of the
filling process.
If the vent valve fails to close and the hissing sound and oxygen vapor
cloud continue, carefully remove the Portable by depressing the release
button on the Reservoir. Venting from the bottom of the Portable will
stop in a few minutes. Allow the unit to warm until you can close the
vent valve. The Portable may require as much as 60 minutes to restore
adequate pressure for accurate oxygen flow. If needed, use an alternate
source of oxygen such as a flow control valve attached to the Reservoir.
6. The Portable does not disengage easily from the Reservoir after filling.
The Portable and Reservoir fill connectors may have become frozen. DO
NOT USE FORCE. Allow a few minutes for the frozen parts to warm,
then disengage the Portable when the ice has melted. To prevent the
units from freezing together, always wipe the male fill connector on the
Reservoir and the female fill connector on the Portable with a clean, dry
cloth before filling.
Cleaning Standard
WARNING: Clean only after the unit is empty and vented.
• Clean using either a household glass cleaner or a solution of mild
dish washing detergent and water.
• Apply cleaning solution directly to a lint-free cloth. Do not spray
cleaners directly on the H300/H850.
• Wipe the outside surface with the lint-free cloth until the outside
surface is clean.
• Do not get cleaner on any internal components or valves.
• Allow the unit to dry thoroughly before using.
HELiOS
TM
WEEE and RoHS
This symbol is to remind the equipment owners to return it
to a recycling facility at the end of its life, per Waste
Electrical and Electronic Equipment (WEEE) Directive.
Our products will comply with the restriction of
Hazardous Substances (RoHS) directive. They will not contain
more than trace amounts of lead or other hazardous material
content.
Transport and Storage
The device should be stored in the upright position well ventilated. Do
not allow the device to be tilted forward or laying on its front. Operating
temperature range of -10°C to 40°C. Storage temperature ranges from
-40°C to 70°C.
Maintenance
Clean the fill connectors on both the stationary and portable units with
a clean, dry, lint-free cloth between each fill to prevent freezing and
possible equipment failure.
NOTE: Any additional maintenance needed must be done by a quali-
fied service technician or service provider.