Programación
[SENSe]:RANGe
{[1]|2|3|4}:AUTO
[SENSe]:RANGe[1]:CAPA
[SENSe]:RANGe[1]:OHM
[SENSe]:RANGe
{[1]|2|3|4} :VOLT
[SENSE]SWEep:OFFSet
:TIME
[SENSe]:VOLTage
{[1]|2|3|4}[:DC]
:RANGe:OFFSet
(Command/Query)
The RANG{[1]|2|3|4}:AUTO <1|0|ON|OFF> command authorizes or prohibits the autoranging
of the selected channel.
ON|1 activates the autoranging.
OFF|0 deactivates this function.
To the question RANG{[1]|2|3|4}:AUTO? the instrument returns the autoranging status for the
selected channel.
(Command/Query)
The RANG:CAPA <range|MAX|MIN|UP|DOWN> command selects the range of measurement
to be used in capacitance mode.
<range> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, the value is expressed in Farad.
To the question RANG:CAPA? the instrument returns the range value of the capacitance.
Response format:
<range><NL>
value in format <NR3>
(Command/Query)
The RANG:OHM <range|MAX|MIN|UP|DOWN> command selects the measurement range to
be used in ohmmeter mode.
<range> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, it is expressed in Ohm ().
To the question RANG:OHM? the instrument returns the value of the measurement range of
the ohmmeter.
Response format:
<range><NL>
value in format <NR3>
(Command/Query)
The RANG{[1]|2|3|4}:VOLT <range|MAX|MIN|UP|DOWN> command selects the
measurement range to be used in voltmeter mode for the selected channel.
<range> is a value in NRf format, it may be followed or not by a multiple and by the unit.
By default, it is expressed in volt.
To the question RANG{[1]|2|3|4}:VOLT? the instrument returns the value of the measurement
range of the voltmeter for the selected channel.
Response format:
<range><NL>
value in format <NR3>
(Command/Query)
The SWE:OFFS:TIME <time|MAX|MIN|UP|DOWN> command sets the horizontal offset of the
trace (run-after-delay or postrig).
<time> is a signed value in format <NRf> ; it may be followed or not by a multiple and by the
unit.
By default, it is expressed in second.
To the question SWE:OFFS:TIME?, the instrument returns the current run-after-delay.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
(Command/Query)
The VOLT{[1]|2|3|4}:RANG:OFFS <offset|MAX|MIN|UP|DOWN> command sets the vertical
offset of the time representation of the selected signal.
<offset> is a value in NRf format, it may be followed or not by a multiple and the unit.
By default the value is expressed in volt.
To the question V{[1]|2|3|4}:RANG:OFFS?, the instrument returns the vertical offset of the
selected signal.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
101