COMPACT PIEZO LED
06.1 Û ELECTROMAGNETIC COMPATIBILITY EN 60601-1-2
WARNING: Contraindications.
Interference with other equipment.
Though compliant with the standard IEC
60601-1-2, the compact piezo LED may
EN
nonetheless interfere with other devices
nearby.
The compact piezo LED must not be used
near to or stacked on other devices. If
adjacent or stacked use is necessary, the
compact piezo LED and the other devices
must be observed and checked to verify
normal operation in the configuration in
which they will be used.
WARNING: Contraindications.
Interference from other equipment. The
use of an electrical scalpel or other electro-
Guidance and manufacturer's declaration - Electromagnetic emissions
The compact piezo LED is intended for use in the electromagnetic environment specified
below. The customer or user of the compact piezo LED should ensure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
14
surgical unit/s near the compact piezo LED
may interfere with its correct functioning.
communication appliances may affect the
correct functioning of the device.
EMC precautions and must be installed
and activated in accordance with the EMC
information given in this paragraph.
producer of the dental chair on which
the compact piezo LED has been installed
is responsible for the compliance of the
product and of the entire system to the
IEC/EN 60601-1 in law.
Compliance
Electromagnetic environment – Guidance
Group 1
The compact piezo LED only uses RF energy for
internal function. Therefore, its FR emissions
are very low and are not likely to cause any
interference with nearby electronic
equipment.
Class B
The compact piezo LED is suitable for use
in all establishments, including domestic
establishments and those directly connected to
Class A
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Complies
WARNING: Portable and mobile radio
WARNING: The device requires specific
WARNING: The technician, or the