2.4 Interruption of programmes

A running programme can be stopped at all times. A programme stop is
triggered by pressing the W - key. "StOP" appears in the display. The
stop must be confirmed within 5 sec by again pressing the W - key,
otherwise the programme continues to run.

2.5 Programming

In addition to the preprogrammed temperatures and heating times,
another 4 temperatures and another 5 heating times can be pro-
grammed.
Prior to programming all active programmes must be terminated. During
programming only the keys on the left are active.
1. Switch off heating block with the main switch.
2. While pressing and holding down the left W - key, switch on the heating
block. Hold down the W - key until a long signal tone is heard and the
left temperature display is flashing.
3. Select a free temperature setting with the X - key. This is indicated
by [ _ _ _° ]. Also a user-defined temperature can be changed.
4. Confirm with the W - key.
5. The first digit flashes. Select desired digit with X - and Y - keys in scroll
mode ( X = digit up, Y = digit down) and confirm with W - key.
6. The second digit flashes, select with X - and Y - keys and confirm with
W - key.
7. The third digit flashes, select with X - and Y - keys and confirm with
W - key. In case of incorrect input the programming cycle can be re-
peated by again pressing the W - key.
8. The X - and Y - keys are used to toggle between time and tempera-
ture.
9. The time is input in the same way as the temperature.
10. The relevant memory location is cleared by inputting X = 000 or
Y = 00:00.
11. Programming is terminated by switching off the instrument.
Programmed temperatures and heating times can be used independ-
ently on both heating units.