Environmental Conditions; Guidance And Manufacturer's Declaration; Electromagnetic Emissions; Electromagnetic Immunity - Invacare SoftCloud Ace Manual Del Usuario

Ocultar thumbs Ver también para SoftCloud Ace:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 61
SoftCloud® Ace, SoftCloud® Pro

9.3 Environmental Conditions

Operation
10 °C – 40 °C
Ambient
temperature
30% – 75%,
Relative humidity
non-condensing
Atmospheric
70 – 106 kPa
pressure

9.4 Guidance and Manufacturer's Declaration

The mattress system is intended for use in the electromagnetic environment specified below. The customer or the user of the
mattress system should ensure that it is used in such an environment.

Electromagnetic Emissions

Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3

Electromagnetic Immunity

Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/ burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
18
Storage and
Transport
-10 °C – 60 °C
10% – 70%,
non-condensing
50 – 106 kPa
Compliance
Electromagnetic environment – guidance
Group 1
The mattress system use RF energy only for their internal functions. Therefore, its
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Class B
The mattress system is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Class A
Complies
IEC 60601 test level
±8 kV contact
±15 kV air
±2 kV for power supply lines
± 1 kV line(s) to line(s)
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Compliance level
±8 kV contact
±15 kV air
±2 kV for power supply lines
±1 kV differential mode
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
70% U
(30% dip in U
)
T
T
for 25 cycles
<5% U
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the mattress system requires
continued operation during power mains
interruptions, it is recommended that
the mattress system be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
1630732-C

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Softcloud pro

Tabla de contenido