Power
frequency
(50/60 Hz)
magnetic
field
IEC61000-
4-8
NOTE U
test level.
The blood pressure monitor is intended for use in the
electromagnetic environment specified below. The customer
or the user of the blood pressure monitor should assure that
it is used in such an environment.
Immunity
Test
16
I
Spengler
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
3 A/m
is the a.c. mains voltage prior to application of the
T
electromagnetic Immunity
IEC
60601
test level
3Vrms
150 kHz
to 80 MHz
3 V/m
80MHz to
2.5GHz
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Compliance
Electromagnetic
Level
Environment
Guidance
Portable and mobile
RF communications
equipment should
be used no closer
to any part of the
blood pressure
monitor, including
cables, than the
recommended
separation distance
calculated from the
3 Vrms
equation applicable
to the frequency of
the transmitter.
recommended
separation
distance
D = 1.2√ P 80 MHz
to 800 MHz
d = 2.3√ P 800 MHz
to 2.5 GHz
Where P is the
maximum output
3 V/m
power rating of the
transmitter in watts
(W) according to
the transmitter
manufacturer and d
is the