Maintenance
The micro motor requires no maintenance. - Check at least every 6 months the cable of the
micro motor. In case of deterioration, it must be replaced at an authorized service center.
Packaging
Before sterilizing the engine, put it in an envelope of Kraft-Polypropylene. The bagging can
maintain the sterility of the material for an extended period of up to 30 days if stored in a dry
and temperature stable environment. Double bagging will increase the time of sterility
maintenance twofold.
Sterilization
Materials: Steam sterilizer according to EN 13060, envelopes of Kraft-Polypropylene.
Parameters: 134°C (273°F) per 5 minutes (2.1-2.2 bars; 30.88 psi).
Shipping and Storage
Temp +5°C/+65°C (+41°F/+149°F) RH < 20-95% non condensing.
conditions
10.1 IRRIGATION LINE MAINTENANCE
The irrigation lines and clips distributed by SATELEC are disposable and must always be discarded after use. Re-
use of an irrigation line or clip may result in contamination of your patients. Do not attempt to resterilize or
modify the irrigation lines and clips.
10.2 CONTRA-ANGLE MAINTENANCE
Refer to the manufacturer's instructions for your contra-angle.
10.3 ROTARY INSTRUMENT MAINTENANCE
Please refer to the manufacturer's instructions for your instruments (burs, drills, etc.).
10.4 DEVICE MAINTENANCE
Important:
Do not use an abrasive product to clean the device.
Do not allow cleaning/disinfecting spray to penetrate inside the device.
The device, the bracket and the multifunction footswitch are not sterilizable.
Unplug the main unit from power source before cleaning or disinfecting. Do not use directly on liquids and
sprays. Never immerse the unit or its parts in liquid.
External surface of the console may be cleaned with a damp cloth or disinfectant wipe. The housing is not
waterproof. Do not use ultrasonic or steam sterilizers.
The I-SURGE control unit, the control footswitch and the bracket must always be cleaned and disinfected after
each procedure using alcohol, disinfectant agents or disinfectant wipes for dental practices.
Regular monitoring of the I-SURGE unit is necessary in order to detect any problem.
10.5 MICRO-MOTOR AND CORD
Do not lubricate the micro motor. Do not attempt to disassemble the motor or motor connector. The saline
solution corrodes all the metal parts. Immediately remove residue.
17