Satelec Acteon I-Surge Manual De Uso página 21

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 56
•Possible interference.
•Broken connections inside.
•Display damaged.
•PCB damaged.
Fault detected: The keyboard is not working properly
Possible cause
•Keyboard damaged.
•Broken connections inside.
Fault detected: The sound signals are not working properly
Possible cause
•Broken connections inside
Fault detected: The motor doesn't work
Possible cause
• Micro motor cable not connected properly Check to see that the micro motor cable is plugged into the connector on the
•Contra-angle defective or not properly
calibrated.
• Footswitch damaged.
•Wrong voltage.
•PCB damaged.
•Motor damaged.
•Connector pins and/or cable damaged.
Fault detected: The motor loses power.
Possible cause
•Contra-angle defective or not properly
calibrated.
•Motor damaged.
Fault detected: The motor stops suddenly
Possible cause
•Micro motor/Cable damaged
•Wrong voltage.
• Footswitch damaged.
Check to see if the central unit has been positioned too close to another electrical
device which can cause interference. In this case, move the interfering electrical
device further away from the central unit.
If you cannot find the cause of the malfunction, immediately discontinue use of
the device and contact SATELEC after-sales service.
Solution
Attempt to change parameters by using the footswitch.
Contact SATELEC after-sales service.
Solution
Immediately discontinue use of the device and contact SATELEC after-sales
service.
Solution
front right of the central unit, disconnect and reconnect.
Check the integrity of the Contra-angle. Remove the Contra-angle and perform
calibration.
Attempt to start the Micro Motor by using the [Motor ON/OFF button] on the
central unit keyboard. If unit does not turn on, contact SATELEC after-sales
service.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
Immediately discontinue use of the device and contact SATELEC after-sales
service.
Solution
Check the integrity of the Contra-angle by removing the Contra-angle and perform
calibration.
Contact SATELEC after-sales service.
Solution
Check to see that the micro motor cable is plugged into the connector on the
front right of the central unit, disconnect and reconnect. Start the motor by
pressing the [Motor ON/OF] button on the keyboard. If the motor turn on,
immediately discontinue use of the device and contact SATELEC after-sales
service.
Verify that the voltage matches the power supply. If it does, try to switch the
central unit off and then on again, or connect the device to another electrical
outlet.
Attempt to start the Micro Motor by using the [Motor ON/OFF button] on the
central unit keyboard.
If unit does not turn on, and contact SATELEC after-sales service.
21

Publicidad

Tabla de contenido
loading

Tabla de contenido