All the pump controls are located on the Turbo-
DRY front panel. The available controls are:
− "POWER" illuminated pushbutton (white lamp):
when pressed the pump is switched on and the
integral lamp lights;
− "LOW SPEED" illuminated pushbutton (green
lamp): when pressed the low speed function of
the pump controller is activated and the integral
lamp lights;
− "Full Speed" LED (green): when this LED is ON
the pump has reached the maximum allowed
speed;
− "Fault" LED (yellow): when this LED is ON a
fault condition has been detected.
The Turbo-DRY technical characteristics are de-
tailed in the following table.
SPECIFICATIONS
TURBO-DRY 65
ISO 63
or CFF
4.5"
flange
Pumping speed N
60 l/s
2
Pumping speed He
30 l/s
Pumping speed H
20 l/s
2
Roughing speed
12 l/min
Base pressure
1.0 x 10
Max power
266 VA at 220 Vac; 290 VA at 110 Vac
Power supply
200 to 240 Vac or 100 to 140 Vac (fac-
tory preset)
Weight
15 kg (35 lbs)
NOTE
No accessories are available for the Turbo pump,
so the Turbo-DRY cannot be used when the heater
band or the water cooling are necessary, as de-
tailed into the Turbo pump technical manual.
STORAGE
When transporting and storing the Turbo-DRY the
following environmental requirements should be
satisfied:
• temperature: from -20 °C to +70 °C
• relative humidity: 0 - 95% (without condensa-
tion)
TURBO-DRY 70
KF 40
ISO 63
KF 40
or CFF
or CFF
or CFF
2.75"
4.5"
2.75"
flange
flange
flange
40 l/s
68 l/s
46 l/s
27 l/s
60 l/s
52 l/s
20 l/s
45 l/s
42 l/s
12 l/min
-6
-9
mbar
2.0 x 10
mbar
17 kg (40 lbs)

PREPARATION FOR INSTALLATION

The Turbo-DRY is supplied in a special protective
packing. If this shows signs of damage which may
have occurred during transport, contact your local
sales office.
When unpacking the Turbo-DRY ensure that it is
not dropped or subject to any form of impact.
Do not dispose of the packing materials in an un-
authorized manner. The material is 100% recycla-
ble and complies with EEC Directive 85/399.
The following figure shows the Turbo-DRY packag-
ing.
INSTALLATION
The Turbo-DRY installation does not require spe-
cific presettings.
An adequate mains voltage connection is neces-
sary to connect the Turbo-DRY power supply.
During operation the following environmental
conditions must be respected:
• maximum pressure: 2 bar
• temperature: from +5 °C to +35 °C
• relative humidity: 0 to 95% (without condensa-
tion).
! ! ! !
The Turbo pump of the Turbo-DRY can reach high
temperature that can cause severe injury.
Pay attention during the unit handling.
46
INSTRUCTIONS FOR USE
WARNING!
87-900-895-01(A)