Descargar Imprimir esta página

Stimulator Programmable Characteristics - Boston Scientific Vercise PC DB-1140 Instrucciones De Uso

Publicidad

Vercise™ PC Implantable Pulse Generator DFU

Stimulator Programmable Characteristics

The stimulation parameters are independent for the two DBS Leads such that stimulation of
two different brain targets can have different amplitudes, pulse widths, stimulation rates, and
contact configurations. The two DBS Leads can also have differing monopolar and multipolar
configurations; however, a given DBS Lead contact cannot be programmed as both monopolar and
multipolar. The programmable parameter ranges for the Stimulator are shown in Table 2.
Note: Some frequency combination may not be used. See the Programming Manual for
specific frequency combinations.
Table 2: Stimulator Function
Parameter
1
Amplitude
2, 3
Rate
4
Pulse Width
Cycle (On/Off)
Stim Ramp On
Contact Connections
Independent Areas of Stim (4
Programs with 4 Areas per Program)
The programmable coverage for each individual contact is limited to 12.7 mA.
1
A programming interlock is enforced to limit the total output current to 20 mA
or less per coverage area. For example, a maximum current output of 12.7 mA
on one contact would limit the total summed current output on the remaining
contacts to 7.3 mA within one coverage area.
The rate is limited to a sum of 255 Hz for a given port.
2
The use of higher frequencies than the ones that are established (250 Hz) is
3
the sole responsibility of the user.
The use of lower pulse widths than established (60-450 μsec) is the sole
4
responsibility of the user.
Vercise™ PC Implantable Pulse Generator DFU
90960114-02 Rev A 12 of 361
Range
0.1 – 20 mA
2 – 255 Hz
20 – 450 µs
1 s – 90 min
1 – 10 s
16
16

Publicidad

loading

Este manual también es adecuado para:

Vercise pc nm-3138-55Vercise pc sc-4252Vercise pc sc-4254