Descargar Imprimir esta página

Connecting The Stimulator - Boston Scientific Vercise PC DB-1140 Instrucciones De Uso

Publicidad

CAUTION: Do not use polypropylene sutures as they may damage the suture sleeve. Do not
suture directly onto the DBS Extension or use a hemostat on the DBS Extension
body. This may damage the DBS Extension insulation.

Connecting the Stimulator

Dual Lead Connection
Connect Left DBS Extension to port C
1. Fully insert the male end of the DBS Extension into the Stimulator until it stops.
a. Insert the header plug to verify no Setscrews obstruct the socket.
b. Wipe the DBS Extension contacts before inserting.
c. Insert the DBS Extensions into the header. When fully inserted, the tip of the DBS
Extension will slide into the back of the port and the retention sleeve on the DBS Extension
will be located under the Setscrew.
CAUTION: Verify proper DBS Extension insertion by checking impedances before tightening
the Setscrew. Tightening the Setscrew down onto a contact can damage the DBS
Extension.
Stimulator Implantation
Connect Right DBS Extension to port D
Vercise™ PC Implantable Pulse Generator DFU
90960114-02 Rev A 5 of 361

Publicidad

loading

Este manual también es adecuado para:

Vercise pc nm-3138-55Vercise pc sc-4252Vercise pc sc-4254