Trouble shooting
Code C6
Number of revolutions per measuring run 5 to 25
revolutions possible, factory–set to 10*
Note
Reducing the number of measurement revolutions will
reduce the accuracy of measurement. The selected
mode of operation can be transferred to the permanent
memory.
Code C8
Selecting the limit (threshold) value for suppression
of minor unbalance readings in grammes, or
ounces. The unit (gms or oz) depends on the
settings made under C3.
grammes:
Range 3.50 to 20.0 g
Factory-adjusted to 3.5 g
Read out limit, e. g. 3.50 g
Select another limit, e. g. 5.50 g
Ounces:
Range 0.25 to 2.00 oz
Factory-adjusted to 0.25 oz
Read out limit, e. g. 0.25 oz
Select another limit, e. g. 0.50 oz
The selected mode of operation can be transferred to
the permanent memory.
Code C10
Storing the selected mode of operation in the
permanent memory
0* =
No storage
1 =
Data are stored in the permanent memory
Store in the permanent memory – a three-tone
signal is given to acknowledge acceptance.
if the mode stored in the permanent memory is to be
changed, enter the desired state (e. g. on or off) for the
mode in question and transfer it via code C10 into the
permanent memory (not possible for code C4).
* = Factory adjusted mode
Operator's Manual
EEWB746A