5.0  CALIBRATION 
Calibration may be required when the balance is initially installed or moved to a new 
location. Over time and use, mechanical deviations may occur so it is recommended 
that you calibrate you balance regularly for peak performance.  
Before calibrating your should make sure you have the correct weights. The weights 
should be known to an accuracy that is appropriate for the balance being calibrated, 
for  example,  OIML  Class  M1  type  or  ASTM  E617  Class  4.  If  you  do  not  have  the 
correct weights do not attempt calibration.  
Model  
CQT 202 
Calibration 
Weight  
100g 
Or 
200g 
 
1) To  calibrate  the  balance  apply  power  and  switch  the  balance  on,  whilst  the 
display is cycling through it's test routine press the [Tare] and  [Unit] keys at 
the same time, then release after 1 second . 
2) The display will show UnLOAD remove any objects from the pan.  
3) Press the [Tare] key to enter a zero value.   
4) The  value  last  used  for  calibration  will  be  displayed  (see  above  table).  To 
select a different calibration weight, press the [Unit] key. 
5) Once  you  have  selected  the  calibration  weight  to  be  used  press  the  [Tare] 
key. 
6) LOAD will be displayed, place the calibration weight on the pan.  
7) Once the stable sign is shown press the [Tare] key.  
8) If  the  weight  is  within  5%  of  the  last  calibration  value  the  display  will  show 
PASS  and  exit  the  calibration  mode.  If  the  calibration  is  not  successful  the 
display will show FAIL and exit the calibration mode, if the calibration fails try 
it again. 
© Adam Equipment 2016
CQT 251  
CQT 601 
200g
100g 
Or 
Or 
400g 
200g 
Or 
600g 
13
CQT 1501 
CQT 2601  
500g
Or 
1000g 
1000g 
Or 
Or 
2000g 
1500g 
308669539revB6-Mar17
CQT 2000  
CQT 5000 
1000g
1000g 
Or 
Or 
3000g 
2000g 
or 
5000g 