Cuando ya se ha escogido un método de control de tempe-
ratura, se visualiza el menú de entrada:
Si quiere mantener 94 ºC durante 5 minutos, presione [9] [4]
[0] y pulse [ENTER] para confirmar. En este momento, el cur-
sor se moverá automáticamente por delante de "m" (de minu-
tos)", presione [5] y [ENTER], el cursor se moverá hasta los
segundos.
Si acepta, pulse [ENTER], en pantalla visualizará:
Importanto: Este seguro de pulsar [CANCEL] para borrar los valores anteriores antes
de introducir uno nuevo, y debe presionar [ENTER] para guardar un nuevo dato.
Nota: Rango disponible de temperatura: 0 ºC a 99.9 ºC. Rango disponible de minutos:
0 min a 99 minRango disponible de segundos: 0 sec a 59 sec
Si quiere seleccionar "Ciclo", mueva el cursor a Cycle.
Enter:
a1
Segment # 2
hold CYCLE end
0 step PCR
Presione [ENTER]
Si quiere un ciclo de 3 pasos, introduzca [3].
Pulse [ENTER]
Step # 1
0.0C Hold 0m 0s
Si quiere mantener 94 ºC durante 30 segundos, presione
[9][4][0] y [ENTER], y presione [3][0] y [ENTER].
Pág. 12
Manual de instrucciones 50604xxx
CASTELLANO
Enter:
a1
Segment # 1
Hold Cycle END
Enter:
a1
Hold at 0.0 C
Hold for 0m 0s
Enter:
a1
Hold at 94.0C
Hold for 5m 0s
Enter:
a1
Segment # 2
hold cycle END
3 step PCR
Step # 1
94.0C Hold 0m 30s
Revisión 1, Abril-08
If you want to get other running information, press [▼] key, and you could see:Press [▼]
key once more to return to the former display.
RUN:
a1
Est remain: 1h01m20s
Step # 1 Tot Cyc 35
0h18m20s Segment # 2
If you want to review the gradient temperature when the instrument is running a gra-
dient step (model 4196), press [▼] key once more and LCD will display as below:
RUN:
a1
Gradient 30C
50C
Gradient View:
Col1 30C left
If the program is finished normally, the LCD will show as below:
Press [CANCEL] return to the main menu
RUN:
Running
finishes at
1h10m0s
Press Key
Pause
At any time in running, the protocol could be paused
by pressing [PAUSE].
Press [PAUSE] again, the protocol will continue. You
will find if the temperature is ramping while you press [PAUSE], the pause function will
not work till the next temperature point is reached.
Stop
At any time in running, the protocol could be stop-
ped by pressing [STOP]. And the screen will ask you:
Press [STOP] again to confirm stop protocol, or press [ENTER] to continue.
Auto Restart
If a power failure occurs when a protocol is running,
the instrument will hold the protocol in memory. When
power is restored, the protocol will begin running
again at the point at which it was stopped, and a notice about the power interruption will
be displayed. The LCD window will show like below:
The message could be cleared by pressing [CANCEL] key, the protocol's runtime scre-
en will immediately be displayed.
Version 1 April-08
Instruction manual 50604xxx
ENGLISH
<<Pause>> 94.0
Timer //// 0h19m40s
Press<STOP>to abort,
<ENTER> to run again
History: Power Lost!
Page 29