5. Wash hands thoroughly again after
handling meter.
6. Verify that the System is working
properly by performing an
Automatic Self-Test (see Getting
Started).
Stop using the meter and call Customer
Care for assistance at 1-800-803-6025,
Monday - Friday, 8AM-8PM EST, if:
• Meter display appears cloudy or any display segments
are missing,
• Markings on meter, including back meter label, are
coming off or are missing,
• Buttons on the meter are hard to push or do not work,
• Unable to insert test strip into Test Port,
• Automatic Self-Test gives an error message.
Note: Other disinfectants have not been tested. The effect
Note: Super Sani-Cloth Wipes have been tested on the
TMX_AIR_OB_GutsENG_REA4TVH03r53.indd 46
TMX_AIR_OB_GutsENG_REA4TVH03r53.indd 46
of other disinfectants used interchangeably has not
been tested with the meter. Use of disinfectants other
than Super Sani-Cloth Wipes may damage meter.
meter for a total of 260 cleaning and disinfecting
cycles, which is equal to cleaning and disinfecting the
meter once a week for a 5 year period. The use life of
the meter is 5 years.