Messages

Display
TMX_AIR_OB_GutsENG_REA4TVH03r53.indd 53
TMX_AIR_OB_GutsENG_REA4TVH03r53.indd 53
Reason
Invalid
Hematocrit
Temperature
Error
Too Cold/
Too Hot
Sample Not
Detected or
Using Wrong
Test Strip
Used Test Strip,
Test Strip outside
of vial too long,
Sample on top of
Test Strip.
Meter
Error
53
Action
Repeat with new test strip,
using capillary whole blood
from the finger or forearm.
If error persists, call
1-800-803-6025, Monday -
Friday, 8AM-8PM EST.
Move meter and test strips
to area between 41°F-104°F;
wait 10 minutes for system
to reach room temperature
before testing.
Retest with new
test strip and larger sample.
Repeat with new test strip.
Make sure Sample Tip of test
strip touched top of sample
drop. If error persists, call
1-800-803-6025, Monday -
Friday, 8AM-8PM EST.
Call 1-800-803-6025,
Monday - Friday, 8AM-8PM
EST.