Table 3
For all me equipment and me systems that are not life-supporting
Guidance and manufacturer's declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
BM47 is intended for use in the electromagnetic environment specified below. The customer or the user of BM47 should assure that it is
used in such an environment.
Immunity test IEC 60601 test
level
3 V
rms
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Compli-
Electromagnetic environment – guidance
ance level
Portable and mobile RF communications equipment should be used no closer to any
part of BM47, including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2
3 V
d = 1,2
80 MHz to 800 MHz
d = 2,3
800 MHz to 2,5 GHz
3 V/m
Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each frequency range.
a
Interference may occur in the vicinity of equipment marked with the following symbol:
58
b