(1)
Observation with and without glasses
When observing without glasses, use the equipment with the eye cup extended.
To do this, turn the eye cup (Fig. 6/1) upwards and to the left (anticlockwise) until it
locks in the highest position.
The eye cup can be locked in four positions – in the upper and lower plus in two
intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing with glasses, turn the eye cup downwards and to the right
(clockwise) until it locks in the lowest position (Fig. 6).
Cleaning and replacing the eyecups
To replace the eyecups or for cleaning, they can be completely screwed off of the
binoculars. Turn the eyecups out as shown in Fig. 6 to the top click stop position
and turn one more thread in the same direction until the eyecup is completely
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the
lowest setting. Finally, you can again set your desired distance between the eye and
eyepiece using the click stops.
Adjusting the interpupilary distance
By folding the halves of the binoculars about the central axis, the eye relief can
be adjusted such that a circular image is formed when observing with both eyes
(Fig. 7).
Note: Depending on the set of the interpupilary distance, the aiming mark and
display may be slightly angled.
R
(3)
(1)
Fig. 6
Focusing the aiming mark and diopter compensation
Use the range finder button (Fig. 1/6) to switch on the aiming mark (Fig. 2/9) and
hold down the button. Focus the aiming mark and the display by turning the right
hand diopter compensation (Fig. 1/4) to the left or the right. Then carefully use
the central focusing (Fig. 1/2) for sharp focusing of the image in the right hand
binocular tube.
Next, use the left hand diopter compensation (Fig. 1/3) for sharp focusing on the
same object of the image in the left hand binocular tube. The values that are set can
be read from the "+" or "–" scale on the back of the binoculars.
Menu settings in the rangefinder
Press the SET button for approx. two seconds to load the menu of your ZEISS Victory
RF. You can then switch between the individual menu options by pressing the SET
button once for each option.
Table 1
Menu 1
Brightness
Menu 2
Ballistics settings
Menu 3
Display settings
Menu 4
Unit of measure
Menu 5
Measuring mode
Menu 6
Key layout
Menu 7
Turn off
Brightness
The brightness settings are located in menu 1 (press the SET button for 2 seconds).
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button,
you can select one of the 11 different brightness curves (curve 1 ≙ darkest setting
possibility; curve 11 ≙ lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment
for each of the selected brightness curves.
Simply release the range finding button (Fig. 1/6) when it reaches the appropriate
setting for you. The brightness curve displayed most recently will then be saved. To
check, you can always display the saved settings in the menu.
Fig. 7
L
(2)
(2)
R
R
(4)
(4)
R
R
Fig. 8
Press SET button for 2 seconds
Press SET button for 2 seconds
+ press the SET button once
Press SET button for 2 seconds
+ press the SET button twice
Press SET button for 2 seconds
+ press the SET button three times
Press SET button for 2 seconds
+ press the SET button four times
Press SET button for 2 seconds
+ press the SET button five times
Press SET button for 2 seconds
+ press the SET button six times
L
L
4 | 7 EN