Troubleshooting
Check the following if the device should fail during operation.
Problem
Nothing happens
when the power switch
is pressed.
No nebulization or
low nebulization rate,
when the power is on.
The compressor is
abnormally loud.
The compressor is
very hot.
Note: If the suggested remedy does not solve the problem, do not try to repair the
device - the device are not user serviceable. Return the device to an authorized
OMRON retail outlet or distributor.
Cause
Is the AC adapter connected
correctly to a power outlet and
the compressor?
Is there medication in the
medication tank?
Is there too much/little
medication in the medication
tank?
Is the vaporizer head missing
or not assembled correctly?
Is the nebulizer kit assembled
correctly?
Is the nozzle blocked?
Is the nebulizer kit tilted at a
sharp angle?
Is the air tube connected
correctly?
Is the air tube folded?
Is the air tube blocked?
Is the air filter dirty?
Is the air filter cover attached
correctly?
Is the compressor being
covered?
Remedy
Check that the AC adapter is connected to
a power outlet and the compressor. Unplug
then reinsert the plug if necessary.
Add the correct amount of medication to
the medication tank.
Attach the vaporizer head correctly.
Assemble the nebulizer kit correctly.
Make sure that the nozzle is free of
blockages.
Make sure that the nebulizer kit is not tilted
at an angle of more than 45 degrees.
Make sure that the air tube is correctly
connected to the compressor and nebulizer
kit.
Make sure that the air tube is not twisted or
bended.
Make sure that the air tube is free of
blockages.
Replace the air filter with a new one.
Attach the air filter cover correctly.
Do not cover the compressor during
operation.