4. Care After Use

• If organic solvents, etc., adhere to the powder case cap, this may result in breakage.
• The powder case cover will deteriorate over time under the influence of ultra violet rays and other
environmental factors and is therefore considered a consumable item. Replace powder case cover
every six months.
• Always thoroughly clean the powder case, the powder case cap and the thread on the powder case
and the powder case cap. Do this once a week and check the powder case cover and powder case
cover for cracks and the scratches.
• Always replace the powder case cap or the powder case immediately should you detect any cracks. If
such a case is used as it is, the powder case cover may break, and this may cause injury by its
fragments.
• In order to prevent the Prophy-Mate neo from clogging put the entire instrument in an ultrasonic bath
once a week. (See 5-1 Ultrasonic cleaning.)
• In the case of failure, make sure to contact your distributor for repairs.
• After use always clean thoroughly internally and externally and ensure all powder residues are
removed and the instrument is totally dry for storage.
• Never lubricate Prophy-Mate neo.(However, lubricate the O-rings with the enclosed turbine oil (See
4-4 Lubrication.))
FLASH pearl is almost insoluble in water. To avoid clogging of the pipes by using FLASH pearl, periodical
cleaning and maintenance of the vacuum hose, saliva ejector and separator are necessary.
4-1 Discard of powder
1) Open the powder case cap, and remove the
remaining cleaning powder. (Fig. 5)
WARNING
CAUTION
NOTICE
8
Fig. 5