ELECTROMAGNETIC IMMUNITY

The EntraFlo™ Feeding Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the EntraFlo™
Feeding Pump should ensure that it is used in such an environment.
Guidance And Manufacturer's Declaration-Electromagnetic Immunity For All Equipment And Systems
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions
and voltage variations on
power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a. c. mains voltage prior to application of the test level.
IEC 60601 Test Level
± 6 kV (contact)
± 6 kV (contact)
± 8 kV (air)
± 8 kV (air)
± 2 kV for power supply lines
Not Applicable
± 1 kV for input/output lines
Not Applicable
± 1 kV differential mode
Not Applicable
± 2 kV common mode
Not Applicable
< 5% UT (>95% dip in UT ) for
Not Applicable
0,5 cycle
40% UT (60% dip in UT ) for
5 cycles
70% UT (30% dip in UT ) for
25 cycles
< 5% UT (>95% dip in UT ) for
5 sec
3 A/m
3 A/m
Compliance Level
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of the
EntraFlo™ Feeding Pump requires continued operation
during power mains interruptions, it is recommended
that the EntraFlo™ Feeding Pump be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commer-
cial or hospital environment.