Indications For Use; Introduction; Cautions - Medline EntraFlo Instrucciones De Operación

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EntraFlo™ Enteral Delivery System - Operating Instructions
Before using the EntraFlo™ Nutrition Delivery
System read this entire manual. It is important
that you understand and are able to follow
all the instructions in this manual. If you have
questions or concerns about any of the instruc-
tions, see back cover to contact your Medline
Sales Representative or Customer Service.

INDICATIONS FOR USE

This pump is intended for the delivery of ente-
ral nutrition and hydration at controlled rates
to patient's gastrointestinal system.

INTRODUCTION

GENERAL
Upon receipt of a new pump, plug it in for a
minimum of 12 hours to charge the backup
battery. For maximum battery life, charge
the battery at least every 30 days, even if
the pump is not in use. Whenever the backup
battery is used to power the pump, charge for
12 hours as soon as conveniently possible. The
backup battery is designed to provide tempora-
ry power to the pump if AC outlet power is
interrupted. Frequent, extended operation of
the pump on the backup battery will shorten
battery life, requiring service by an authorized
service representative. The EntraFlo™ Nutrition
Feeding Pump is an easy-to-use, versatile
device for the appropriate administration of
enteral feeding solutions.
KEY FEATURES OF THIS DEVICE INCLUDE:
Ease of Use
Versatility
Rate adjustable in 1 mL/hr increments from 1
to 295. Dose adjustable in 5 mL increments
from 0 - 9995.
Lightweight and Portable
An internal backup battery can provide pow-
er for short-term use. While the pump should
normally be operated on AC outlet power,
the internal backup battery can provide
operation for up to 8 hours at 100 mL/hr if
necessary.
Ease of Maintenance
The fire resistant plastic case and sealed
control panel helps prevent fluid leakage and
allow cleaning with any mild soap.
Alarms
Low Battery, Battery On, Dose Complete,
and Occlusion/Empty for detecting empty
container or occluded line.
2 ½ minute Pause Alarm
Low Occlusion Pressure
15 p.s.i. occlusion allows for safe use with any
enteral feeding tube.
Built-in Pole Clamp
Total Volume Delivered Display
Accumulated Volume Display*
• Adjustable Alarm Volume
• Dose Limit for Intermittent Feedings
With 3 selectable alarm modes.*
• Memory Retention of Infusion Rate, Dose
Limit, Volume Delivered and Accumulated
Volume even after unit is turned off.
Settings are automatically recalled when the
pump is turned back on.
• Serviceability
Self-diagnostic mode allows for complete
circuitry check with the push of two buttons.
*with serial numbers greater than 400000

CAUTIONS

PUMP
• NOT FOR PARENTERAL USE INTENDED
EXCLUSIVELY FOR ENTERAL FEEDING
• Unauthorized modification to any
component of the pump may adversely affect
performance or safety in unpredictable ways.
• Turning rotor may pinch fingers.
• Plug and power outlet should be clear of
obstruction.
• To disconnect pump from AC power pull plug
from power outlet.
• Operator should be within a distance that
allows visibility of readouts and ability to hear
alarms.
• Wireless communications equipment, such as
mobile phones, can affect this device and
should be kept at least 4 inches (0.1 meter)
from the device during operation.
• Do not operate pump where there is a lot of
dust, dirt or airborne particles.
• Do not operate pump with flow sensor in
direct path of intense light.
• If necessary to dispose of pump, do so in
EN
1

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ent199235

Tabla de contenido