2. MACHINE CLASSIFICATION
• Type and Degree of protection against electric shock:
Class II/Type BF
Degree of protection against water (vertically falling drops) stipulated
in IEC 60529 : IPX 0.
Machine should not be used at the place close to fl ammable Anes-
thetics and Oxygen.
Operation Mode: intermittent load in consecutive operation
Loading Time: max. 3 minutes
Resting Time: min. 10 minutes.
3. PRODUCT FEATURES
• Simple interface design for easy control
• Freedom of activity during operation by wireless handpiece
• Ergonomic and compact body
<Endo mode>
• RPM : (10:1) 200~900RPM(900rpm±10%)
(16:1) 100~600RPM(600rpm±10%)
(20:1) 100~600RPM(600rpm±10%)
• Torque : (10:1) 1~2N·cm (16:1) (20:1) 1~4N·Cm
• Gear Ratio : 10:1, 16:1, 20:1,
• Auto Reverse Mode : Level 1, 2
<Program function>
• 5 memory programs of
(Gear Ratio, Torque, RPM, Direction mode) are available to set
• Programs no. 6 and 7 are reciprocating modes and the user can set
desired angle for forward or reverse rotation
<Setting torque limit function>
• If actual load on the fi le is higher than preset torque, it stops work-
ing automatically.
<Auto reverse mode>
• Users can choose Auto- mode according to operation condition
Level 1: Once actual torque is over the preset torque, the fi le
move in Reverse rotation. After removing the load, it
stops.
Level 2: Once actual torque is over the preset torque, after
auto-reverse and load released, it turns forward-rota-
tion again.
<Auto-OFF function>
• Power is off with "Beep" sound after 4 minutes without operation.
(During the motor operation even without use, power does not
switch off automatically)
6
BAE370R_IFU-Final.indd 6
Machine has been confi rmed that it complies with the rules of medical
machinery IEC 60601-1 : 2007 which provides reasonable protec-
tions against harmful interference in typical medical installations.
Machine can radiate radio frequency energy. If not installed or used in
accordance with the instructions, it might cause harmful interference
possibly to other devices in nearby, however, it is not assured that it
will occur to a particulate installations.
If machine cause harmful interference to other devices and that inter-
ference is determined by their ON/OFF, users are encouraged to try
to correct the interference by one or more of the following measures:
- Machine needs to be used by a medical specialist or qualifi ed person.
- Do not use machine at the place too close to other devices.
- Consult with manufacturer or qualifi ed service technicians for the
electronic safety information of the machine.