2. INSTALLATION

2.1 Select the Location

Avoid heat sources, rapid temperature changes, air current or excessive vibrations. Allow sufficient
space.
2.2 Leveling
Be sure the balance is level before it is used or after its
location is changed.
The PX balance has a level bubble in a small round
window beside the display.
To level the balance, adjust the 4 leveling feet until the
bubble is centered in the circle.
Please refer to the right figure for leveling.
2.3 Connecting Power and Acclimatising the Balance
Connect the DC output connector to the power receptacle on the rear of the balance. Then connect
the AC adapter plug to a suitable electrical outlet.
Acclimatising
It is suggested that the balance should not be used until it has been connected to power and
acclimatised to the environment for a certain period of time. In the case of a balance with the
precision above 0.1 mg, the acclimatisation time should be 1.5 hours; in the case of balance with the
precision of 0.01 mg, the acclimatisation time should be more than 4 hours.

2.4 Connecting the Interface

The PX balance has two data interfaces, RS232 and USB.
Use the RS-232 port to connect either to a computer or a printer with a standard (straight-through)
serial cable. Use the USB port to connect to a computer with a USB 2.0 Type A to Type B cable.
Interface connections on the rear of the balance
USB
RS232