- optical sensor
Transmission wavelength: 890 nm.
Luminous power transmitted: depends on sighting distance ;
at 1cm =>
0,5mW/cm²
at 50cm =>
2mW/cm
Minimum luminous power received: 10µW/cm2.
Ratio of reflecting area to target area: > 5%.
Detection distance: from 1 to 50cm.
Sighting angle from the perpendicular to the target: 0° ± 15°.
6.4 CHARATERISTICS OF THE ADAPTER AND ITS END FITTINGS
- Mechanical adapter
End fittings: elastomer, Shore hardness 80
Pressure exerted on the moving part: between 2 and 40N.
Maximum speed: 10,000rpm.
Life: approximately 1,000 hours at 3,000rpm at a pressure of 20N.
- Conical end fitting accessory
This end fitting is used for a measurement by contact on the shaft end of a system in rotation.
It is an elastomer cone (max. diameter 15mm) that fits onto the output shaft of the adapter, with quick locking.
Minimum measurement shaft diameter: 5mm.
- Cylindrical end fitting accessory
This end fitting is used for a measurement by contact on the shaft end of a system in rotation.
It is an elastomer cylinder that fits onto the output shaft of the adapter, with quick locking.
It is used to measure the speeds of shafts larger than 5mm in diameter or having flat ends.
- End fitting accessory with wheel
This end fitting is used to measure a linear speed by direct contact with the moving part.
It is a rigid elastomer wheel that fits onto the output shaft of the adapter, with quick locking.
Diameter of the wheel: 30.183mm.
Circumference of the wheel: 10cm ± 0.1mm.
.
56