Product Overview

The SIP 10 Sipper module is an optional accessory for the DR 3900
spectrophotometer. The SIP 10 serves to improve measurement
accuracy. The same optical characteristics apply to both the zero
compensation and the measurements. All test solutions flow through a
cell and any errors due to optical differences between redundant cells
are eliminated. A peristaltic pump pumps a constant amount of fluid
through the Pour-Thru Cell.
Scope of Delivery
Please check that the order is complete. If anything is missing or
damaged, please contact the manufacturer or distributor immediately.
SIP 10 Sipper module, pump tubing pre-assembled
Dual pathlength Pour-Thru Cell, 1 inch/1 cm
USB connection cable
®
Tygon
tubing, 1.70 m (5.6 ft)
Stainless steel tube adapter
User manual, multilingual
Installation
N O T I C E
Make sure that there are no kinks in the various tubes.
Check the fill level of the waste vessel on a regular basis.
N O T I C E
There may always be residual fluid in the tubes. Always wear suitable protective
clothing and goggles to change the tubes or perform maintenance work in order to
prevent contact with the skin or eyes.
N O T I C E
Connect the device directly to the photometer via the USB cable.
Do not use a USB hub connection.
1.
Position the SIP 10 to the left of the photometer.
2.
Clean the Pour-Thru Cell with a lint-free cloth.
3.
Insert the Pour-Thru Cell into the cell compartment of the
photometer in the direction indicated by the arrow.
4.
Cut a piece of tubing no longer than 40 cm (1.3 ft) to serve as the
feed tube.
5.
Connect one end of the feed tube to the IN connection of the Pour-
Thru Cell.
6.
Connect the other end of the feed tube to the stainless steel tube
adapter and place this in the zero-solution/sample vessel.
7.
Cut a piece of tubing approx. 30 cm (1 ft) long to serve as the
connection tube.
8.
Connect the connection tube to the OUT connection of the Pour-
Thru Cell and the input connection of the peristaltic pump.
9.
Use the remaining tubing as a discharge tube, and connect one
end to the output connection of the peristaltic pump.
10. Place the other end of the discharge tube in a suitable waste
vessel.
11. Connect the SIP 10 USB cable to a USB port on the photometer.