Problem
Display error message:
Check tube!!
Key if done
Display error message:
SYSTEM ERROR
Call service!
Display error message:
High leak in
last session.
Possible Cause
The air tubing is loose or blocked.
Component failure.
You have experienced excessively
high leak levels during the night.
Solution
Check that the air tubing is
connected securely to your mask
and the air outlet on the front of
the AutoSet Spirit.
When you have checked the air
tubing, press the Front key to
restart the machine.
If this does not clear the message,
turn the power switch off and then
turn it back on to restart the
machine.
Return your AutoSet Spirit for
servicing.
Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask properly.
If this message appears again,
contact your clinician.
T
ROUBLESHOOTING
31