Maintenance - DeVilbiss 7305 Serie Guía De Instrucciones

Ocultar thumbs Ver también para 7305 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4.
For multi-patient use: After washing:
a.
After parts are completely dry, place jar and lid in autoclave with the open end down. Ensure parts are not touching. Run one sterilization steam cycle at 250°F
(121°C) for 15 minutes. NOTE-Jar is guaranteed up to 30 cycles of autoclave sterilization at the indicated conditions.
b. Dispose of and replace filter, tubing and elbow between patients.
Suction Unit (single-patient use)
1.
With the power switch in the "Off" position, disconnect the DeVilbiss Suction Unit from all external power sources.
2.
Wipe the housing with a clean cloth and any commercial (bacterial-germicidal) disinfectant.
CAUTION– Do not submerge in water as this will result in damage to the vacuum pump.
NOTE–Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean the unit.
tubing (single-patient use)
1.
Disconnect the tubing from the unit.
2.
Rinse thoroughly by running warm tap water through it.
3. Follow by soaking in a solution of 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) for 60 minutes. Rinse with clean, warm water
and air dry.
4.
Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
carrying case (single-patient use)
1.
Wipe the case using a clean cloth dampened with detergent and/or disinfectant.
NOTE– Disinfection information is based on AARC Clinical Practice Guideline Suctioning of the Patient in the Home.
Suction Unit (multi-patient use)
Device Cleaning and Disinfection When There is a Patient Change
When medical devices have already been used with a patient, contamination with human pathogenic germs should be assumed (unless there is evidence to the contrary), and
the next patient, user or third party should be protected by appropriate handling and preparation. Therefore, when there is a patient change, people must be protected during
the transport and handling of the device and the device must be fully processed, i.e., cleaned and disinfected, by suitably trained personnel before reuse to protect the next
patient. The complete processing may only be done by the manufacturer or by a qualified DeVilbiss provider/service technician.
NOTE – When the unit is used as per instructions under normal conditions the interior of the unit is protected from exposure to pathogens by the in-line filter on the collection
container, hence no disinfection of internal components is necessary.
NOTE – If the unit is used without an in-line filter then the interior of the unit has been exposed to pathogens and the unit cannot be disinfected.
NOTE – If the following processing of the unit by a qualified DeVilbiss provider/technician is not possible, the unit must not be used by another patient!
DeVilbiss Healthcare recommends that at least the following procedures be carried out by the manufacturer or a qualified third party between uses by different patients.
1. D ispose of all accessory components that are not suitable for reuse, i.e., collection container, filter, tubing and carrying case.
2. With the power switch in the "Off" position, disconnect the DeVilbiss Suction Unit from all external power sources.
3. Visually inspect unit for any damage, missing parts, etc.
4. Wipe the housing with a clean cloth and a commercial (bacterial-germicidal) disinfectant that meets the requirements listed in the NOTE below and is used as per the
disinfectant manufacturer's recommended dilution rates and instructions.
CAUTION – Do not submerge in water as this will result in damage to the vacuum pump.
NOTE – Do not use any cleaners or disinfectants that contain ammonia, benzene and/or acetone to clean the unit.

maINtENaNcE

Inspect suction tubing and container for leaks, cracks, etc. before each use.
DaNGER
Electric shock hazard. Do not attempt to open or remove cabinet, there are no user-serviceable internal components. If service is required, return unit to a
qualified DeVilbiss Healthcare provider or an authorized service center. Opening or tampering with the unit will void warranty.
changing Filter cartridge (single-patient use) Disposable container
1. Change filter cartridge if overflow occurs or every two months, whichever comes first.
Turn unit
"off".
2.
3. Remove filter cartridge and 43/8" tubing.
4.
Install new cartridge and tubing.
NOTE– Do not substitute any other material for this filter cartridge. Substitution may lead to contamination or poor performance; use only DeVilbiss filter cartridges.
NOTE– The filter cartridge contains a hydrophobic filter. If the filter media becomes wet, air flow will be stopped. The filter cartridge must then be replaced. Do not remove filter
media from filter cartridge.
NOTE– Filter cartridges are included with each disposable container. They are also available separately (7305D-635 12/pack).
changing bacteria Filter (single-patient use) Reusable container and/or Disposable container
1. Change bacteria filter if overflow occurs or every two months, whichever comes first.
2. Remove filter by disconnecting it from suction unit and lid assembly.
3. Replace with a clean DeVilbiss bacteria filter (non-sterile) and remount to suction unit and lid. Ensure that the clear side of the bacteria filter is toward elbow and
container when installing/re-installing. Do not reverse direction of filter. Additional filters (7305D-608 12/pack) may be purchased from your authorized DeVilbiss
Healthcare provider.
NOTE– Do not substitute any other material for this bacteria filter. Substitution may lead to contamination or poor performance; use only DeVilbiss filters.
NOTE– Bacteria filter must be changed between patients.
A-704
EN - 7

Publicidad

Tabla de contenido
loading

Tabla de contenido