At a fixed flow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure
setting, patient breathing pattern, mask selection, and leak rate. This warning applies to most types of CPAP
devices.
Never allow the oxygen source to run continuously while connected to the IntelliPAP if the device is not in use. If the
IntelliPAP is not being used, turn off the oxygen flow.
CAUTION
Oxygen is a prescribed gas and should only be administered under the supervision of a physician.
The flow setting of the oxygen source must be specified by a physician.
Supplemental oxygen is sometimes prescribed in addition to CPAP for treatment of Obstructive Sleep Apnea. If your
physician prescribed oxygen for your sleep therapy, you can add the oxygen in one of two ways:
1.
Obtain the optional oxygen adapter (part #7353D-601) from your homecare provider and connect it to the air supply
port on the IntelliPAP. Connect one end of the air supply tubing to the open end of the adapter and the other end of
the air supply tubing to the mask.
2.
If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your
mask.

ACCESSORIES/REPLACEMENT ITEMS

Replacement Items
Air-inlet filter (4/pk)
Fine particle filter (4/pk)
Air supply tubing (6')
Air supply plug
Humidifier Chamber
Power Items
AC power cord (USA)
AC power cord (UK)
AC power cord (EU)

MAINTENANCE

DANGER
Electric shock hazard-Do not attempt to open or remove the cabinet; there are no user-serviceable components
inside. If service is required, contact your equipment provider for instructions on obtaining service. Opening or
attempting to service your device will void the warranty.
WARNING
Disconnect supply before servicing.
Pressure Accuracy–The IntelliPAP is designed to provide accurate pressure performance from the factory without
additional calibration. No routine service is required provided the device is used in accordance with the manufacturer's
directions.
NOTE–Some countries require periodic servicing and calibration of this kind of medical device. Contact your equipment
provider for further information.
Standard Air-Inlet Filter–Check the filter every 10 days and clean it if necessary. Replace the filter every 6 months or
sooner if it is damaged. The standard filter is designed to filter particles larger than 3.0 microns in size. The IntelliPAP
MUST have the standard filter in place during operation.
Optional Fine Particle Filter–Check the filter every 10 days and replace it if it is dirty or damaged; otherwise replace the
filter every 30 days. The optional fine particle filter is designed to filter particles as small as 0.3 microns in size.
EN - 10
DV51D-602
Heater connection cover
DV51D-603
Oxygen adapter
CPAP-TUB 6
Carrying case
DV51D-604
Heated Humidifier
DV5C
Chamber Sealing Gasket
DV51D-606
AC power cord (Australia)
DV51D-608
DC power cord
DC battery clamp-on adapter (requires DC
DV51D-607
Power Cord DV51D-619)
DV51D-605
7353D-601
DV51D-610
DV5HH
DV5C-614
DV51D-609
DV51D-619
DV51D-696
A-DV53