Rugby 260SG / 270SG / 280DG, Setup
Head speeds
Scan and 0 rps
Lock function: Protects the Rugby against unwanted changes.
If no button is pressed, the display will revert to the main screen after ten seconds
and stores possible changes.
Rugby 260SG: fixed to 10 rps (600 rpm).
Rugby 270SG: 5 or 10 rps (300 or 600 rpm).
Rugby 280DG: 0, 2, 5 or 10 rps (0, 120, 300, 600 rpm) and additionally has the
scan features scan 10, 45 or 90 degrees.
Select the head speed that provides the best performance for your application. The
selected head speed is shown in the status line.
The scan feature and head speed 0 rps is only available on a Rugby 280DG.
Scan can be selected in the same screen as head speeds.
Use the UP or DOWN buttons to select scan 10, 45 or 90 degrees.
Once selected the head will stop and scan in the selected area.
From the main menu press the DOWN button to rotate the head clockwise. Press UP
to rotate counter-clockwise.
Pressing UP and DOWN at the same time will rotate the head in 90 degree incre-
ments.