Multipatient use

WARNING
An antibacterial filter is mandatory if the device is used on multiple patients.
In a mulitpatient use environment, you must perform the following before the device is
provided to a new patient:
Air filter and the
antibacterial filter
Mask
Air tubing
Device
Humidifier

Servicing

CAUTION
Inspection and repair should only be performed by an authorized agent. Under no
circumstances should you attempt to open, service or repair the device yourself.
This product should be inspected by an authorized ResMed service center five years from
the date of manufacture, except for the internal battery which must be replaced every two
years from the manufacturing date of the device. Prior to this, the device is intended to
provide safe and reliable operation provided that it is operated and maintained in
accordance with the instructions provided by ResMed. Applicable ResMed warranty
details are provided with the device at the time of original supply. Of course, as with all
electrical devices, if any irregularity becomes apparent, you should exercise caution and
have the device inspected by an authorized ResMed service center.

Troubleshooting

If there is a problem, try the following suggestions. If a problem cannot be solved, contact
ResMed.

Alarm troubleshooting

The most common reason for an alarm to sound is because the system has not been
properly assembled. Check that the air tubing has been properly attached to the device and
patient interface (and humidifier if used).
Notes:
• The alarm log and alarm settings are maintained when the device is powered down and
in the event of a power loss.
• If multiple alarms are active simultaneously, the alarm with the highest priority will be
displayed first.
• If an alarm activates repeatedly, discontinue use and return the device for servicing.
Replace.
Reprocess; Cleaning, disinfection and sterilization instructions
are available from the ResMed website, www.resmed.com/
masks/sterilization. If you do not have Internet access, please
contact your ResMed representative.
Replace the air tubing. Alternatively, consult the air tubing
instructions for cleaning and disinfection information.
Disinfect the Stellar as follows:
1. Apply undiluted mikrozid
clean non-dyed disposable cloth.
2. Wipe all surfaces of the device, including the air outlet;
avoid liquid entering any openings in the device.
3. Leave the disinfectant to work for five minutes.
4. Wipe residual disinfectant from the device with a clean,
dry non-dyed disposable cloth.
As instructions for humidifiers vary, see the user guide for
the humidifier in use.
®
®
AF or CaviCide
liquid to a
Troubleshooting
21