Appendix A: EMC Information
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
Flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD) IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
Compliance
Electromagnetic Environment-Guidance
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
Group 1
likely to cause any interference in nearby electronic
equipment.
Class B
The device is suitable for use in all establishments,
Class A
including domestic establishments and those directly
connected to the public low-voltage power supply
network.
Complies
IEC60601
test
level
±6kV contact
±8kV air
±2kV for power
supply line
±1kV for input/out
line
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
<5 % U
(>95 % dip
T
in U
)for 0,5 cycle
T
40 % U
(60 % dip
T
in U
)for 5 cycles
T
Electromagnetic
Compliance
Environment-Guidance
Floors should be wood,
concrete or ceramic tile. If
±6kV contact
floors are covered with
synthetic material, the relative
±8kV air
humidity should be at least
30%.
±2kV for power
Mains power quality should be
supply line
that of a typical commercial or
±1kV for input/out
hospital environment.
line
± 1 kV line(s) to
Mains power quality should be
line(s)
that of a typical commercial or
hospital environment.
<5 % U
(>95 % dip
T
Mains power quality should be
in U
) for 0,5 cycle
T
that of a typical commercial or
40 % U
(60 % dip
hospital environment. If the
T
in U
) for 5 cycles
user of this device requires
T
15