• Flow rate (ml/h), volume (ml) and interval (h) are displayed through the large characters. Words below
describe what the number relates to (ml/h, total volume, volume bolus, interval bolus, total volume del).
The pump also displays messages, for example "END OF DOSE" will appear when the pump has finished
delivering a single feed dose.
Figure 6
The back light of the display turns off 10 seconds after the last key is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
• The Flocare
Infinity
®
• Insert the pump set in the pump, switch the pump on and fill the pump set when necessary (see paragraph:
filling the pump set).
• The display shows the total volume delivered since the pump was last cleared.
• If necessary clear the volume delivered by pressing the "clr" key.
• The last installed flow rate (ml/h) is now displayed.
• The pump is now in the "hold" mode and is ready for programming.
• We strongly recommend that for those patients where interruptions or delay in therapy could affect their health status,
the DOSE setting of the Infinity pump must be used and programmed.
ml/h
• Adjust the flow rate (ml/h) if required by using the "+" or "-" key. Hold either key down to change rapidly.
• If another parameter is indicated in the LCD, it will be necessary to first press the "ml/h" key followed by the
"+" or "-" key to set the flow rate.
• The flow rate ranges from 1 - 400 ml/h in 1ml increments.
• The flow rate slows down and stops shortly at 50 - 125 - 250 ml/h.
• Press the "volume ml" key and make sure the value is disabled by pressing the "clr" key. The display will
indicate the word "CONT".
• Press the "bolus ml/int h" key and make sure the volume bolus value is disabled by pressing the "clr" key.
The display will indicate the word "NO". This means the pump will run in a continuous mode.
• Start the pump by pressing "start/stop".
• The arcs around the word "run" start rotating.
• The flow rate (ml/h) is displayed whilst functioning.
ZMNL900617_AW.indd 82
• The arcs around the run symbol rotate when the pump is
• The remaining battery capacity (see paragraph: battery
• The wall plug symbol indicates that the charger is plugged
PROGRAMMING THE PUMP
+ pump can be used for continuous or dose feeding.
™
CONTINUOUS FEEDING
+
start/stop
–
1 ml/h
running.
operation).
in.
-82-
UK-IE