"LITE ON" (+) the light of the LCD remains on when the pump is connected to the mains. "LITE OFF" (-) the
light of the LCD switches off after a few seconds even if connected to the mains. Use the "+" or "-" key to change
the setting from "LITE ON" to "LITE OFF" or vice versa.
+
OUTP PMDS
fill set
–
OUTP NRSE
Hold the '+' key down and press "ON/OFF" for 2 seconds. The pump enters the set up mode. Press the
"FILL SET" key to switch to the next setting:
"OUTP PDMS" (+) to connect with a Patient Data Monitoring System. "OUTP NRSE" to connect with a NURSE
CALL system. Use the "+" or "-" key to change to setting from "OUTP PDMS", to "OUTP NRSE" or "OUTP
OFF" and vice versa.
Press the
key and keep it pressed for 2 seconds to exit the SET UP mode and turn off the pump. The
settings are automatically saved.