3. Acquiring PANO Images
Laser Beam Aligning
1.
The Vertical Beam is fixed. Align the center of the patient's face (Mid-sagittal Line)
with the Vertical Beam. (It's to prevent horizontal expansion of the image)
2.
Align the Horizontal Beam in a straight line to the Frankfurt Line on the patient's face.
Use the Horizontal Beam lever on the column (left side of the Control Panel) to
position it. Make sure that the Horizontal Beam is aligned to the patient's face
horizontally.
24
Be careful not to shine the laser beam directly into the person's eyes.
Doing so may result in vision loss.
If the laser beam is not correctly positioned, there may be distortion,
causing the image to be enlarged or reduced, or ghost shadows may
occur and lower the image quality. Be sure to align the laser beam
properly.
 Three Laser Beams (Vertical Beam, Horizontal Beam and Canine
Teath Beam) will be activated when CONFIRM button is clicked.
- The Laser Beams are turned off automatically after 20 minutes or
when READY button is clicked.)
 To turn the Laser beams on/off manually, click the
Control Panel of the Handle Frame or the
Console Software.
icon on the
icon on the