Limpieza y esterilización de los insertos reusables
Ocultar thumbs Ver también para PIEZOSURGERY:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

02 Û PRE-CLEANING
Soak the inserts for 10 minutes at 40°C
±2°C. This procedure reduces the quantity
of blood, protein, and mucous present in the
insert;
EN
During immersion into the enzymatic
solution, delicately brush the surface of
the insert with a brush with soft nylon
bristles to eliminate all traces of visible dirt.
Accurately clean the difficult areas such as
sharp edges and especially the interstices
between the cutting cusps;
Thoroughly rinse and scrub the insert with
a with soft bristled nylon brush under
running tap water;
By using a disposable syringe inject and
aspirate the enzymatic detergent solution
into hard-to-reach areas (lumen/cannulae);
Repeat this operation three times for
effective removal of soil from the inner
lumen surface;
Flush the internal channel of the insert with
distilled water injected at a pressure of 3.8
bars for at least 10 seconds so as to remove
any residues;
4
3
4
5
6
7
3,8 bar
10 sec
x3