Gima CONTEC08A Manual De Instrucciones página 70

Ocultar thumbs Ver también para CONTEC08A:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Precaution for SpO
2
Note
• Make sure the nail covers the light. The probe wire should be on the backside of the hand.
• SpO
value always displays in the fixed place.
2
• Testee can not use enamel or other makeup.
• Testee's fingernail can not be too long.
• The fingers which are too cold or too thin may affect the measure accuracy, please insert the thicker
finger such as thumb or middle finger deeply enough into the probe.
• The SpO
probe is suitable for children above four years old and adults(Weight should be between
2
15Kg to 110Kg). The device may not work for all patients. If you are unable to achieve stable readings,
discontinue use it. Do not use the device on infant or neonatal patients.
• The update period of data is less than 5 seconds, which is changeable according to different pulse rate
values.
• Pulse rate waveform through the normalization, when pulse rate waveform becomes smooth and
stable, the read value is optimal value, and the waveform at the moment is the most standard one.
• The maximum temperature for the contact surface of the device with the body is less than 41°C,
and the temperature is measured by a temperature measuring device.
• The SpO
probe has been calibrated before leaving factory.
2
Warning
• Pulse oximeter can overestimate the SpO
chemicals.
• Cables of ES (Electrosurgery) equipment and SpO
• Do not put the SpO
ENGLISH
measurement
probe on extremities with arterial catheter or venous syringe.
2
value in the presence of Hb-CO, Met-Hb or dye dilution
2
probe must not be tangled up.
2
70

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

32902

Tabla de contenido