Cleaning And Sterilization - NSK Perio-Mate Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 67
7-4 Lubrication
1) Dry the O-ring section of the Powder Case and the Adjustment Ring section, apply a drop of the
supplied turbine oil and rub it in with your finger, etc. (Fig. 21, 22)
2) Perform autoclave sterilization.
If you fail to lubricate the O-ring section on the Handpiece side, the Joint section will be
damaged and this may cause the Handpiece connection to be difficult to rotate. It may
CAUTION
also cause water leakage.
After autoclave sterilization, the Flow Volume Adjustment Ring can be difficult to rotate
in some rare cases. However, it may return to the normal state once you rotate it. If the
Adjustment Ring cannot rotate properly, contact your Authorized NSK Dealer.
8

Cleaning and Sterilization

Only the handpiece can be washed via Thermo Disinfector.
Do not use the Thermo Disinfector for the Powder Case, Powder Case Cap and Nozzle.
CAUTION
8-1 Ultrasonic Cleaning
Make sure to perform Ultrasonic Cleaning once a week or after the remedy described in
10. Troubleshooting to prevent clogging of powder.
CAUTION
1) After disassembling, perform cleaning. (See 7-1 Discard of Perio Mate Powder - 7-3 Cleaning and
Washing.)
2) Pass through the supplied Ring Wrench from the Nozzle end, loosen the Nozzle, and remove the
Nozzle from the Handpiece.
3) Perform Ultrasonic Cleaning with hot water for the Handpiece, Powder Case and Nozzle.
4) After cleaning, blow waterdrops from all parts by air to dry them.
5) Reassemble the parts following the disassembly procedures in reverse.
6) Connect the Powder Case and Hose with the cover not attached and discharge air. This allows water
in the Internal Nozzle to be discharged. Then thoroughly dry the inside.
OIL
Fig. 21
14
OIL
OIL
Fig. 22

Publicidad

Tabla de contenido
loading

Tabla de contenido