If excess cleaning powder is put into the Powder Chamber, the correct amount of
powder may not be sprayed immediately after the start of use.
CAUTION
The Perio Mate Powder has the property of easily scattering because of its fine particles.
Perform filling slowly and be careful to prevent the powder from scattering.
Do not use any powder stored for a long while in opened packages. Do not use leftover
powder. Using such powder may cause clogging of the nozzle.
Use only Perio Mate Powder. Other powders have different particle sizes and will clog
and not perform as expected. Using other powders may also cause significant damage
to the Nozzle Tip.
Always securely close the container lid of the Perio Mate Powder. Failure to do so will
attract moisture which may cause clogging.
Use Perio Mate Powder as cleaning powder. If other powder is used, the Perio Mate
Nozzle Tip will be damaged and it will increase the risk of damaging substantia ossea
and dentine or causing emphysema. NSK shall assume no liability for any failures
caused by the use of other powders.
If powder adheres to the Powder Case and/or the threaded portion of Powder Case Cap,
clean it immediately or it will become difficult to close the Cap.
6-6 Operation
1) Attach the Nozzle Tip to the Nozzle of the Perio
Mate Handpiece.
2) After filling the powder, spray it into a spittoon,
etc., for one second and confirm that powder
is mixed with water and air, and spraying is
appropriately performed.
3) Insert the Nozzle Tip into a periodontal pocket
by 3 mm or more. (Fig. 6)
If the Nozzle Tip is inserted less than 3 mm,
powder may not be sprayed into the periodontal
pocket efficiently.
4) Start operation. Spray powder in the same
periodontal pocket for up to 5 seconds or on
the same tooth for up to 20 seconds. Do not
spray longer than this specified time.
5) After treatment, remove the Nozzle Tip from
the Perio Mate Handpiece using the Nozzle Tip
Remover and dispose of it. (Fig. 7)
Nozzle Tip Remover
8
Nozzle Tip
Fig. 5
Fig. 6
Fig. 7