7
Troubleshooting
Warning
Temperature
too high. Out of
operating range.
See Owner's
Booklet.
What it means
Meter is too hot (above 111°F) to
work correctly.
What to do
Move the meter and test strips
to a cooler area. Insert a new
test strip when the meter
and test strips are within the
operating range (43–111°F).
If you do not get another
Temperature too high message,
you can proceed with testing.