koban KP 2102 Manual Del Usuario página 40

Pinza amperimétrica digital
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

KP 2102
Digital clamp meter
Measuring Frequency by V
range:
WARNING
Beware of Electrocution.
Pay special attention to avoid
electric shock when measuring
high voltage.
Do not input the voltage which
more than 750V rms AC.
Plug the black test lead into the
COM jack and the red test lead into
the INPUT jack.
Set the rotary selector to the V∽
range position.
Press the "Hz/%" to switch to
frequency measurement.
Connect test leads to the two ends of
the source or load for measurement.
Take the reading on the LCD.
NOTE:
1.
Frequency test range is 10Hz
-10kHz. It is possible to test the
frequency which is higher than
10kHz but the tolerance of the
test result can not be ensure.
" means the maximum input
2.
"
voltage is 750V rms AC.
Instructions manual |
40
Measuring Frequency by
HZ/DUTY range::
WARNING
Beware of Electrocution.
Pay special attention to avoid
electric shock when measuring
high voltage.
Do not input the voltage which
more than 250V rms AC.
Plug the black test lead into the
COM jack and the red test lead into
the INPUT jack.
Set the rotary selector to the
HZ/DUTY range position.
Connect test leads to the two ends of
the source or load for measurement.
Take the reading on the LCD.
OFF
400A
600 CA T lll
MAX 400A
MS2108A
40A
V
V
SEL
Hz/DUTY
MIN
/%
B.L
REL
Hz
MAX
HOLD
Press 2 Sec
AUTO-POWER OFF
AC/DC CLAMP METER
COM
INPUT
600V
CAT. III
MAX 1000V
750V
C o r r e c t
www.grupotemper.com
OFF
400A
600 CA T lll
MAX 400A
MS2108A
40A
V
V
SEL
Hz/DUTY
MIN
/%
B.L
REL
Hz
MAX
HOLD
Press 2 Sec
AUTO-POWER OFF
AC/DC CLAMP METER
COM
INPUT
600V
CAT. III
MAX 1000V
750V
In c o r r e c t

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido