SECTION 1—GENERAL GUIDELINES
The use of this device is limited to the oxygen patient. Cylinders MUST be
used only by the oxygen patient and are not to be distributed to any other
individual for any purpose.
This product is NOT to be used by patients who breathe through their
mouths.
DO NOT allow oil from your hands or other sources to come into contact
with the compressor coupling or bottle fittings. These substances may
become flammable in the presence of oxygen and cause injury. Avoid
touching the bottle fitting whenever possible. Example of items to avoid
are hand creams or lotions, cooking oils, suntan oils or sunblocks and
similar oily products. If you use these types of products Invacare suggests
you wash your hands prior to using the HomeFill unit. If the bottle fitting
should come in contact with such substances it should be cleaned off with
a dry cloth prior to connecting it to the compressor.
NEVER oil or lubricate the compressor coupler or bottle fittings. NEVER
use any penetrating oil or lubricant such as WD-40 or 3 in One. If
connection difficulty is experienced, verify that the bottle fitting is being
inserted straight down into the compressor coupler. If continued difficulty
is experienced, contact your supplier for assistance.
DO NOT allow cylinders to tip or fall. Secure gas cylinders so they cannot
fall. Ensure cylinder is always in upright position.
DO NOT store cylinders near sources of heat or flame.
This product is equipped with a relief valve. If you hear a loud hissing or
popping sound coming from the product, discontinue use of product, turn
the flow selector to the "0" position, and contact your dealer/Home Care
equipment provider.
This product is designed to operate with a standard or high flow single
lumen, adult oxygen nasal cannula with a maximum length of 2.1 m (7 ft).
The cannula is for single patient use only.
Cannula MUST be installed on outlet connection prior to use.
DO NOT place the unit under clothing while in use. When this product is
in use a small amount of oxygen is vented. Wearing the unit under
clothing may saturate fabrics with oxygen and cause them to burn rapidly
if exposed to sparks or flame. It may take several hours for oxygen levels
in fabrics to return to normal.
DO NOT use HomeFill compressor unit, to fill cylinder outdoors or
anywhere below 13°C (55°F).
NEVER attempt to fill cylinder with anything other than the HomeFill
compressor.
Part No 1167426