16
Instructions for Use
Error Messages
Message
Fault: short
circuit
Overvoltage
Flow meter
alarm
204/260
Description
In normal operation (after the
starting phase) the output
connection is shorted (output
current higher than
6 A).
A failure has occurred in the
controller power supply section,
or the controller has received a
spurious signal.
Alarm relating to the purge flow.
The flow value remained at less
than the set threshold for 10
seconds or more.
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)
Repair action
Check connections and
shortages between pump and
controller. Press the START
push-button twice to start the
pump.
Press the START button
twice to start the pump again.
If the message is redisplayed,
contact Agilent for
maintenance.
Check the correct operation
of the gas supply circuit.