Fuse Holder and Voltage Changer Assembly
The following figure shows the location of this assembly.
Figure 6
Proceed as follows to replace one or both fuses:
Figure 7
To change the power supply voltage rating after having removed the
assembly as explained above, extract the voltage changer and then
reposition it to view the desired voltage rating.
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)
Rear panel
Remove the assembly by levering in position A with a small
screwdriver.
Replace the fuse
Use only T-type fuses of the following characteristics:
- 100/120 Vac
- 220/240 Vac
Voltage Changer
Fuse Holder and Voltage Changer Assembly
3.15 A
6.3 A
Technical Information
211/260
17