17
Technical Information
Examples
WIN
206
207
208
209
210
211
212
213
214
215
300
301
302
400
402
WIN = Window
R = Read
W = Write
T = Type
L = Logical
N = Numeric
A = Alphanumeric
230/260
R
W
T
Description
X
N
ERROR CODE (0÷7) [NO ERROR, OVERVOLTAGE, SHORT
CIRCUIT,CHECK CONNECTION, TOO HIGH LOAD,
OVERRIDE, PUMP OVERTEMP, CONTROLLER OVERTEMP]
X
L
STATE S1 [0=OFF / 1=ON]
X
L
STATE S2 [0=OFF / 1=ON]
X
N
UPPER BEARING TEMPERATURE [0-99 °C]
X
N
LOWER BEARING TEMPERATURE [0-99 °C]
X
N
FLOW READ [0-100 sccm]
X
L
STATUS OUTPUT FLOW METER [0=OFF / 1=ON]
X
N
FLOW METER ALARM TIME RELATING TO THE LAST
EVENT
X
N
FLOW METER ALARM CYCLE NUMBER RELATING TO THE
LAST EVENT
X
N
FLOW METER ALARM EVENT NUMBER
X
N
CYCLE TIME (0-999.999) IN MINUTES
X
N
CYCLE NUMBER (0- 65.535) IN COUNTS
X
N
PUMP LIFE (0-999.999) IN HOURS
X
A
CRC PROGRAM LISTING
X
A
CRC PARAMETER LISTING
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)