Use a soft dry cloth with not – abrasive and not – solvent detergents.
BEFORE STRATING CLEANING OPERATIONS, MAKE SURE THAT INTERNAL PARTS OF DEVICE DON'T COME INTO
CONTACT WITH LIQUIDS
The ASPIMED 2.2 suction equipment does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. Unpack the instrument and always check integrity of plastic parts and
feeding cable, they might have been damaged during previous use. Connect the cable to electrical network and turn the switch on.
Close the aspirator outlet with your finger and with suction regulator at maximum check that the vacuum indicators reaches at least -
75kPa (-0.75 bar). Rotate the knob from right to left. The vacuum indicator should go down -25kPa (-0.25 bar).
Check that no loud noises are present. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and situated in the plug protects
the instrument. For fuses replacing, always the type and the range.
Type of fault
1. The suction unit doesn't work
2. No aspiration
3. No aspiration
4. The Vacuum power on the
patient side is either very low or
absent
5. The float doesn't close
6. The float doesn't close
7. Low suction
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7
If the overfill security system fails, continued aspiration will be stopped by the antibacterial filter.
Should this back-up security system also fail, there is a risk of penetration of liquid into the device. Don't attempt to proceed with
aspiration under these circumstances.
Return the device to MORETTI SpA for service.
The Manufacturer will provide upon request electric diagrams, components list, descriptions, setting instructions and any other
information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT MORETTI
TECHNICAL SERVICE.
MORETTI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO
BE TAMPERED.
CLEANING DEVICE
PERIODICAL MAINTENANCE CHECKS
Cause
Cable is damaged
External power source failure
Jar Cap not screwed on correctly
Lid seal not in its seat
• Vacuum regulator set to minimum
• Protection filter blocked or
damaged
• Connection tubes blocked, kinked
or disconnected
• Shut-off valve blocked or damaged
• Pump motor damaged
If the cap has been washed, ensure
that the float is not partially detached
The float it's covered by dirty material
Foam inside the jar
None of the procedures have
achieved the desired results
Remedy
Replace the cable
Check the external power source
Unscrew the cap, and re-screw it
Unscrew the cap and insert the seal properly in its
seat
• Turn the vacuum regulator clockwise and check
the value of the vacuum on the gauge
• Replace the filter
• Replace or reconnect the tubes, check the jar
connections
• Empty the jar, or disconnect the tube from the
jar and unblock the shut-off valve. The unit twill
only work in the upright position
• Refer to authorised service personnel
Fit the float into it's place
Unscrewed the cap, leave the and put in on
autoclave
Fill the jar to 1/3 full of ordinary water
Contact MORETTI SpA customer service