Description Of Back Panel - SOPRO COMEG 284 Manual De Utilización

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 40
E N G L I S H
7

DESCRIPTION OF BACK PANEL

7.1 DESCRIPTION OF VENTS
The vents [A] located on the back panel of the device must not be obstructed as this can cause a risk of
overheating.
The light source has an automatic safety device which stops the lighting function if the internal temperature is
too high. A pictogram indicating whether this safety system is activated will be displayed on the screen [L2].
7.2 TRANSFORMER / MAINS SOCKET
The electrical supply to the device is via an IEC standard mains cord provided with the equipment.
7.3 MAINS SOCKET – FUSES
The electrical supply to the device is via the power socket located on the back panel [C3], which must
be connected to the mains via the mains cord provided. This socket comprises a fuse flap [F] as well as
the main 'on' switch [S3].
When replacing a fuse, it is essential to disconnect the product from the mains and use the same type of
fuse. The T in 'T800mA' means 'time delay'. Only use UL/CSA labelled fuses.
7.4 INDICATORS AND SYMBOLS ON BACK PANEL
The indicators and symbols on the back panel of the product identify it as conforming to the international
standards IEC 60601-1, IEC 60601-2-18, IEC 60417 and EN 980 (see section 13)..